Model |
P3251 |
Option |
Pressure type |
Negative or positive gauge pressure |
absolute pressure |
Negative or positive gauge pressure |
Output signal |
4 . . . 20 mA - 2-wire system 0 . . . 5 VDC - 3-wire system 0 . . . 10 VDC - 3-wire system |
other signals on request |
Accuracy % of F. S. 1) |
0,5 |
0,25 |
0,5 |
0,25 |
0,5 |
0,25 |
|
Ranges accord. to EN |
0 . . . 0.1 bar 2) to 0 . . . 16 bar
|
0 . . . 25 bar to 0 . . . 600 bar
|
0 . . . 0,25 bar to 0 . . . 16 bar |
|
Sensor element |
piezoresistive |
thin film |
piezoresistive |
|
Repeatability |
¡Â¡¾ 0.05% of F. S. |
|
Stability (annual) |
¡Â¡¾ 0.2% of F. S. in rated conditions |
|
Case |
Stainless steel |
|
Pressure connection |
¡Â 0¡¦1,6 bar G 1 B;¡Ã0...2,5 bar G 1/2 B |
|
Wetted parts |
Stainless steel |
|
Overload limit |
¡Â16 bar 3,5 x; ¡Â600 bar 2 x; |
|
Electrical connection |
plug according to DIN EN 175301-803 form A with junction box round connector M12x1; 4-pin
|
cable outlet with 1 m cable |
Power supply |
10 . . . 30 VDC (14 . . . 30 VDC for output 0 . . . 10 V) |
|
Power consumption |
output 4 . . . 20 mA: signal currency voltage output: 8 mA |
|
Load |
¡Â UB-12V/0.020A for output (0) 4 . . . 20 mA > 5 kOhm for output 0 . . . 5 V > 10 kOhm for output 0 . . . 10 V |
|
Temp. compens. range |
0 . . . 80 ¡ÆC |
|
Temperature influence - Zero point - Measuring range |
¡¾ 0.2% / 10 K 3) ¡¾ 0.2% / 10 K |
|
Adjustability |
zero point and full scale up to ¡¾ 10% |
|
Response time |
¡Â1 ms (within 10% to 90% of F. S.) |
|
Protection type |
IP 65 to EN 60 529 / IEC 529 |
IP 67 for cable outlet |
Emission 4) |
according to EN 61 326 |
|
Interference 4) |
according to EN 61 326 |
|
Electrical protection types |
polarity, overload and short-circuit protection |
|
Temperature ranges - Storage - Medium - Ambient |
-40 . . .100 ¡ÆC -30 . . .100 ¡ÆC -20 . . . 80 ¡ÆC |
medium temperature -40 . . . 125 ¡ÆC
integrated cooling element for temperatures up to 150¡Æ C |
Weight |
approx. 0.2 kg |
|